Entirely automated quantification of human immunodeficiency virus type 1 (HIV-1) RNA in plasma by using the ultrasensitive COBAS AMPLICOR HIV-1 monitor test and RNA purification on the MagNA pure LC instrument.
J Clin Microbiol. 2003; 41(3):1248-1251
Doi: 10.1128/JCM.41.3.1248-1251.2003
[OPEN ACCESS]
Web of Science
PubMed
FullText
FullText_MUG
- Führende Autor*innen der Med Uni Graz
-
Berg Jörg
- Altmetrics:
- Dimensions Citations:
- Plum Analytics:
- Scite (citation analytics):
- Abstract:
- The ultrasensitive COBAS AMPLICOR HIV-1 Monitor test was complemented with automated RNA purification on the MagNA Pure LC instrument. This enabled entirely automated ultrasensitive assessment of viral loads in human immunodeficiency virus type 1 (HIV-1)-infected individuals. The detection limit of the fully automated assay and the viral load measurements in 80 clinical samples were found to be in good agreement with those of the conventional ultrasensitive COBAS AMPLICOR HIV-1 Monitor test. The fully automated assay showed markedly reduced hands-on time and was found to be suitable for the routine assessment of HIV-1 viral loads in a clinical diagnostic laboratory.
- Find related publications in this database (using NLM MeSH Indexing)
-
Automation -
-
HIV-1 - physiology
-
Humans -
-
RNA, Viral - analysis
-
Reagent Kits, Diagnostic -
-
Sensitivity and Specificity -
-
Viral Load -