Medizinische Universität Graz Austria/Österreich - Forschungsportal - Medical University of Graz

Logo MUG-Forschungsportal

Gewählte Publikation:

SHR Neuro Krebs Kardio Lipid Stoffw Microb

Saraswat, A; Simionato, L; Dawson, JA; Thio, M; Kamlin, CO; Owen, L; Schmölzer, G; Davis, PG.
Determining the best method of Nellcor pulse oximeter sensor application in neonates.
Acta Paediatr. 2012; 101(5):484-487 Doi: 10.1111/j.1651-2227.2011.02571.x
Web of Science PubMed FullText FullText_MUG

 

Co-Autor*innen der Med Uni Graz
Schmölzer Georg
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
Aim: To identify the optimal sensor application method that gave the quickest display of accurate heart rate (HR) data using the Nellcor OxiMax N-600x pulse oximeter (PO). Methods: Stable infants who were monitored with an electrocardiograph were included. Three sensor application techniques were studied: (i) sensor connected to cable, then applied to infant; (ii) sensor connected to cable, applied to investigator's finger, and then to infant; (iii) sensor applied to infant, then connected to cable. The order of techniques tested was randomized for each infant. Time taken to apply the PO sensor, to display data and to display accurate data (HRPO = HRECG +/- 3 bpm) were recorded using a stopwatch. Results: Forty infants were studied [mean (SD) birthweight, 1455 (872) g; gestational age, 31 (4) weeks; post- menstrual age, 34 (4) weeks]. Method 3 acquired any data significantly faster than methods 1 (p = 0.013; CI,) 9.6 to) 3.0 sec) and 2 (p = 0.004; CI,) 5.9 to) 1.2 sec). Method 3 acquired accurate data significantly faster than method 1 (p = 0.016; CI,) 9.4 to) 1.0 sec), but not method 2 (p = 0.28). Conclusion: Applying the sensor to the infant before connecting it to the cable yields the fastest acquisition of accurate HR data from the Nellcor PO.
Find related publications in this database (using NLM MeSH Indexing)
Equipment Design -
Heart Rate -
Humans -
Infant, Newborn -
Oximetry - instrumentation
Reproducibility of Results -

Find related publications in this database (Keywords)
Infant
newborn
Monitoring
Neonatal intensive and special care
Pulse oximeter
© Med Uni Graz Impressum