Nardon, E; Glavač, D; Benhattar, J; Groenen, PJ; Höfler, G; Höfler, H; Jung, A; Keller, G; Kirchner, T; Lessi, F; Ligtenberg, MJ; Mazzanti, CM; Winter, G; Stanta, G.
A multicenter study to validate the reproducibility of MSI testing with a panel of 5 quasimonomorphic mononucleotide repeats.
Diagn Mol Pathol. 2010; 19(4):236-242 Doi: 10.1097/PDM.0b013e3181db67af
Web of Science PubMed FullText FullText_MUG Google Scholar

 

Co-authors Med Uni Graz
Höfler Gerald
Winter Gerlinde
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
Microsatellite instability (MSI) testing in clinics is becoming increasingly widespread; therefore, there is an urgent need for methodology standardization and the availability of quality control. This study is aimed to assess the interlaboratory reproducibility of MSI testing in archive samples by using a panel of 5 recently introduced, mononucleotide repeats (MNR). The quality control involved 8 European institutions. Participants were supplied with DNA extracted from 15 archive colon carcinoma samples and from the corresponding normal tissues. Every group was asked to assess the MSI status of the samples by using the BAT25, BAT26, NR21, NR24, and NR27 mononucleotide markers. Four institutions repeated the analysis using the NCI reference panel to confirm the results obtained with the MNR markers. The overall concordance among institutions for MSI analyses at single locus level was 97.7% when using the MNR panel and 95.0% with the NCI one. The laboratories obtained a full agreement in scoring the MSI status of each patient sample, both using the mononucleotide and the NCI marker sets. With the NCI marker set, however, concordance was lowered to 85.7% when considering the MSI-Low phenotype. Concordance between the 2 panels in scoring the MSI status of each sample was complete if no discrimination was made between MSI-Stable and MSI-L, whereas it dropped to 76.7% if MSI-L was considered. In conclusion, the use of the MNR panel seems to be a robust approach that yields a very high level of reproducibility. The results obtained with the 5 MNR are diagnostically consistent with those obtained by the use of the NCI markers, except for the MSI-Low phenotype.
Find related publications in this database (using NLM MeSH Indexing)
Carcinoma - diagnosis
Carcinoma - pathology
Colon - pathology
Colonic Neoplasms - diagnosis
Colonic Neoplasms - pathology
DNA - genetics
DNA - isolation & purification
Humans -
Microsatellite Instability -
Molecular Diagnostic Techniques - methods
Molecular Diagnostic Techniques - standards
Pathology, Molecular - methods
Pathology, Molecular - standards
Reproducibility of Results -

Find related publications in this database (Keywords)
quality control
microsatellite instability
mononucleotide repeats
formalin-fixed
paraffin-embedded tissues