Eggenstein, C; Borchardt, M; Diekmann, C; Gründig, B; Dumschat, C; Cammann, K; Knoll, M; Spener, F.
A disposable biosensor for urea determination in blood based on an ammonium-sensitive transducer.
Biosens Bioelectron. 1999; 14(1):33-41 Doi: 10.1016/S0956-5663(98)00103-1
Web of Science PubMed FullText FullText_MUG

 

Führende Autor*innen der Med Uni Graz
Spener Friedrich
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
A potentiometric urea-sensitive biosensor using a NH4(+)-sensitive disposable electrode in double matrix membrane (DMM) technology as transducer is described. The ion-sensitive polymer matrix membrane was formed in the presence of an additional electrochemical inert filter paper matrix to improve the reproducibility in sensor production. The electrodes were prepared from one-side silver-coated filter paper, which is encapsulated for insulation by a heat-sealing film. A defined volume of the NH4(+)-sensitive polymer matrix membrane cocktail was deposited on this filter paper. To obtain the urea-biosensor a layer of urease was cast onto the ion-sensitive membrane. Poly (carbamoylsulfonate) hydrogel, produced from a hydrophilic polyurethane prepolymer blocked with bisulfite, served as immobilisation material. The disposable urea sensitive electrode was combined with a disposable Ag/AgCl reference electrode to obtain the disposable urea biosensor. The sensor responded rapidly and in a stable manner to changes in urea concentrations between 7.2 x 10(-5) and 2.1 x 10(-2)mol/l. The detection limit was 2 x 10(-5) mol/l urea and the slope in the linear range 52 mV/decade. By taking into consideration the influence of the interfering K(+)- and Na(+)-ions the sensor can be used for the determination of urea in human blood and serum samples (diluted or undiluted). A good correlation was found with the data obtained by the spectrophotometric routine method.
Find related publications in this database (using NLM MeSH Indexing)
Biosensing Techniques -
Humans -
Membranes, Artificial -
Potentiometry -
Quaternary Ammonium Compounds -
Transducers -
Urea - blood

Find related publications in this database (Keywords)
disposable biosensor
disposable reference electrode
potentiometric biosensor
urea biosensor
blood urea